LI-600 Porometer/Fluorometer Instruction Manual
8-6
from the handle, guiding the battery connector out
through the connector slot in the instrument head.
7
Insert the new battery.
Gently insert the new battery into the handle, guid-
ing the battery connector in through the connector
slot in the instrument head. Place the wires behind
the tab.
8
Replace all parts in reverse order.
Replace the dust cap, endcap, and instrument handle
coverings. Plug in the battery connector and replace
the instrument head shells.
After a battery replacement or any other repair, verify
the safe state of the instrument by checking that the dis-
play is and the fan are powered off.
Replacing the air intake filter
The air intake filter serves to remove solid particles from
air before it is drawn through the LI-600 porometry sys-
tem. A 0.45 micron filter (part number 301-19838) is a
direct replacement for the original 1.0 micron filter
(part number 301-16918).
Follow these steps to replace the air intake filter.
1
Remove the shells.
Remove the shells on the instrument head. Set the
shells safely aside (see
2
Remove the filter cover.