Section 8. Maintenance
8-13
Figure 8-11
. Attach a tank of ultrapure dry air to the cal-
ibration clamp.
Figure 8-12
. Visual representation of an attached air tank
7
Dry the RH sensors.
Deliver an airstream across the RH sensors for 30
minutes to 1 hour. Do not allow maximum airflow
to exceed 0.25 LPM. The Flow or inlet flow sensor
will read negative values (i.e., approximately -30 to -
100 µmol s
-1
) due to the airflow toward both the air
exhaust and air inlet. Flow_s will read positive values
(i.e., approximately 100 to 200 µmol s
-1
).
RH sensor zero
Follow these steps when zeroing the RH sensors.