Section 9. Theory of operation
9-15
9-42
where
P
width
is 667E-9 seconds
The total amount of light incident on the leaf from the
LEDs is then calculated from
9-43
Freq
set
will be either the steady-state modulation fre-
quency
Mod
rate
if not during a flash,
Freq
flash
during a
rectangular flash or Phases 1 and 3 of an MPF or the cal-
culated frequency based on ramp rate during Phase 2 of
an MPF.
PAR
flr
is written to flash files during each log
point.
Dark-adapted vs light-adapted measurements
With the LI-600 fluorometer, the actinic light source is
provided by external sources unknown to the instru-
ment. The instrument cannot know if a given meas-
urement being made is on a dark-adapted leaf or not.
The user must decide which type of measurement is
being made.
on the next page shows the values
being used for each configuration type. The calculated
yield uses the same equation in either case, as shown
here.
9-44
9-45