LI-600 Porometer/Fluorometer Instruction Manual
8-2
Regularly check that the IRT sensor is clean. Use
foam swabs and isopropyl alcohol (IPA) to carefully
clean the IRT sensor.
l
Clean the fluorometer.
Regularly check that the fluorometer is clean. Use
foam swabs, a microfiber cloth, and IPA to carefully
clean the fluorometer optics.
l
Clean the RH sensors.
Regularly check that the RH sensors are clean, for
example after a campaign, before extended storage,
or if a shift in RH zero is noted (see
page 8-8). Use a squeeze bottle with deionized (DI)
water to carefully rinse the RH sensors. Do not touch
the active side of the sensors with your fingers or a
swab. Do not use IPA. Dry them with compressed
air, then bake them for 6 hours at 100 to 120 °C.
Never touch the RH sensors or H
2
O buffer material
in the instrument head, as oils on the skin and
scratches can affect the instrument’s performance.
Figure 8-1
. The RH sensors board.
l
Clean the barcode window, quantum sensor,
aperture clamp, and leaf gasket.
Regularly check that the barcode window, quantum
sensor, aperture clamp, and leaf gasket are clean, as