Section 7. Troubleshooting
7-5
as a piece of printer paper, and flush the instrument’s
system until the sample RH value drops to match the
reference RH (RH Ref).
Instrument warnings
This section describes the warnings that can appear on
the LI-600 display and how to address them.
Instrument warnings will appear with this
symbol on the instrument display.
To determine the warning type, open
Instrument Set-
tings
by pressing the left arrow from the Select Con-
figuration screen, then select
Warnings
.
Matching. Do not clamp on leaf.
Matching happens automatically when the LI-600 is
powered on, and when you select Match on the Instru-
ment Settings screen. Always allow the instrument to
match before clamping onto a leaf.
Low memory
The Low Memory warning appears then there is less
than 10% of data storage space remaining. Stop using
the instrument; download the data, then clear all data
from the instrument.
Low battery
The Low Battery warning appears when there is less
than 10% of battery charge remaining. Stop using the
instrument and charge the battery.
Blower off
The Blower Off warning appears when the blower is off.
To turn the blower on, power on the LI-600 and select a