Section 7. Troubleshooting
7-7
Duplicate Flash ID
This is caused by a user taking measurements, manually
changing the clock back an hour from Daylight Saving
Time to Daylight Standard Time, then continuing to
take measurements that result in a duplicate Flash ID
(time stamp).
Unpaired Flash ID Detected
This error occurs when the fluorometer option is
enabled "on the fly." The user picks a configuration
with the fluorometer disabled and takes measurements.
With the instrument connected to the computer, the
user enables the fluorometer option and re-uploads the
configuration to the same spot on the instrument and
continues taking measurements.
Orphaned Flash ID Detected
This error occurs when the fluorometer option is dis-
abled "on the fly." The user picks a configuration with
the fluorometer enabled, takes measurements, then con-
nects the instrument to the computer. The user then dis-
ables the fluorometer option, and uploads the updated
configuration to the same spot on the instrument and
takes more measurements.
Unrecognized Row Type
This error occurs when a barcode with a newline char-
acter is scanned.
Malformed Flash Data Reading
If this error repeatedly pops up, each time referencing a
different row, then the problem is likely a bad USB
cable.