LI-600 Porometer/Fluorometer Instruction Manual
8-12
Connect the LI-600 to a computer via USB then open
the computer software. Click the drop-down menu
on the Toolbar then select Instrument Calibration.
3
Power off the pump.
Select RH Zero then click
Begin RH Zero Cal-
ibration
to view the RH Ref and RH Sample. Click
TURN PUMP OFF
.
4
Remove the leaf clamp.
Gently but firmly pull on the leaf clamp to remove it
from the instrument.
Figure 8-10
. The leaf clamp.
5
Attach the calibration clamp.
Gently but firmly attach the calibration clamp. The
calibration clamp is equipped with a 10-32 hose barb
for tube connection (P/N 9960-327).
6
Attach the air tank.
Connect a tank of ultrapure dry air to the calibration
clamp, ensuring that the tube connection does not
prevent the clamp from sealing properly.