3.18 Control elements (supplement)
4
4. Control panel 1
Menu button
The menu symbol itself (4) remains the same. Howe-
ver, the sequence of indication of the individual
values has been changed.
By pressing the menu button, the different parame-
ters are displayed in the following order (press
menu button once per parameter):
1. Chamber temperature
2. Specimen temperature
3. Preselected value of preset counter
(Press to activate / deactivate preset coun-
ter)
4. Maximum (= lowest) temperature of specimen
head
5. Time (real time)
6. Defrost time (= time when defrost cycle starts)
7. Duration of defrost cycle (Chamber)
5. Control panel 2
/ - Tasten
When the instrument is in sectioning mode, (indica-
tion TRIM --> OFF!) display indication will go to
sectioning thickness selection by pressing or
.
Displayed is always the section thickness value
which has been selected last.
Press or to select a new section thickness
value.
Selectable section thickness selection range:
0
-
2 µm
in 0.5-µm
steps,
2
-
10 µm
in 1-µm
steps,
10
-
20 µm
in 2-µm
steps,
20
-
60 µm
in 5-µm
steps,
60
- 100 µm
in 10-µm
steps,
100
- 300 µm
in 50-µm
steps.
Summary of Contents for CM3050 S
Page 1: ...Service Manual Leica CM3050 S Cryostat...
Page 49: ...Chapter 2 Service Manual Leica CM3050 S Page 11 Version 1 0 12 00 Leica Microsystems...
Page 92: ...Chapter 2 Service Manual Leica CM3050 S Page 54 Version 1 0 12 00 Leica Microsystems...
Page 168: ...Chapter 3 Service Manual Leica CM3050 S Page 76 Version 1 0 12 00 Leica Microsystems...
Page 197: ...Chapter 5 Service Manual Leica CM3050 S Page 5 Version 1 0 12 00 Leica Microsystems...
Page 199: ...Chapter 5 Service Manual Leica CM3050 S Page 7 Version 1 0 12 00 Leica Microsystems...
Page 211: ...Chapter 5 Service Manual Leica CM3050 S Page 19 Version 1 0 12 00 Leica Microsystems...
Page 286: ...Chapter 5 Service Manual Leica CM3050 S Page 94 Version 1 0 12 00 Leica Microsystems...