1.3.31 Technical Data
Operating temperature range (ambient
temperature): + 18°C to + 40°C. All
specifications related to tempe-
rature are valid only up to an ambient
temperature of + 22°C and relative air
humidity of less than 60%!
Refrigeration system
50 Hz
60 Hz
Cryochamber
Temperature selection range
0° C to -40°C (+ 2 K / - 0 K)
0° C to -40°C (+2 K / - 0 K)
Defrosting
Automatic hot-gas defrost cycle,
Automatic hot-gas defrost cycle,
programmable in 15-min increments;
programmable in 15-min. increments;
1 automatic defrost cycle per 24 hours.
1 automatic defrost cycle per 24 hours.
Duration: 6 - 12 minutes
Duration: 6 - 12 minutes
Manual defrost cycle
Manual defrost cycle
Refrigeration capacity
690 W
690 W
Safety factor
3
3
Refrigerant
280 g ±5g refrigerant R 404A
*
280 g ±5g refrigerant R 404A
*
Compressor oil
0.6 l EMKARATE RL-22S, ICI
*
0.6 l EMKARATE RL-22S, ICI
*
Quick-freeze shelf
Maximum temperature
- 43°C (+ 0 K / - 2 K)
- 43°C (+ 0 K / - 2 K)
Number of quick-freeze stations 10
10
Specimen cooling (Option)
Temperature range
-10° C to -50°C ±3K
-10° C to -50°C ±3K
Defrosting
Manual defrost cycle (electric)
Manual defrost cycle (electric)
Refrigeration capacity
320 W
320 W
Safety factor
3
3
Refrigerant
185 g ±5g refrigerant R 404A
*
185 g ±5g refrigerant R 404A
*
Compressor oil
0.4 l alpha 22, Kyodo
*
0.4 l alpha 22, Kyodo
*
or RENISO E22, Fuchs
*
or RENISO E22, Fuchs
*
Type
CM3050S-10
CM3050S-1
CM3050S-8
CM3050S-9
Marks of conformity
-
-
-
-
Nominal voltage
100 V AC ±10% 230 V AC ±10% 240 V AC ±10% 100 V AC ±10%
Nominal frequency
50 Hz
50 Hz
50 Hz
60 Hz
Power draw
1800 VA
1800 VA
1800 VA
1800 VA
Max. start-up current for 5 sec
35 A eff.
25 A eff.
25 A eff.
30 A eff.
Protective fuse
IIII
Automatic cutout
T15A M3
T10A T1
T10A T1
T15A M3
Pollution degree
2
2
2
2
Overvoltage installation category II
II
II
II
Heat emission (max.)
1800 J/s
1800 J/s
1800 J/s
1800 J/s
Summary of Contents for CM3050 S
Page 1: ...Service Manual Leica CM3050 S Cryostat...
Page 49: ...Chapter 2 Service Manual Leica CM3050 S Page 11 Version 1 0 12 00 Leica Microsystems...
Page 92: ...Chapter 2 Service Manual Leica CM3050 S Page 54 Version 1 0 12 00 Leica Microsystems...
Page 168: ...Chapter 3 Service Manual Leica CM3050 S Page 76 Version 1 0 12 00 Leica Microsystems...
Page 197: ...Chapter 5 Service Manual Leica CM3050 S Page 5 Version 1 0 12 00 Leica Microsystems...
Page 199: ...Chapter 5 Service Manual Leica CM3050 S Page 7 Version 1 0 12 00 Leica Microsystems...
Page 211: ...Chapter 5 Service Manual Leica CM3050 S Page 19 Version 1 0 12 00 Leica Microsystems...
Page 286: ...Chapter 5 Service Manual Leica CM3050 S Page 94 Version 1 0 12 00 Leica Microsystems...