60
Version 1.7, Revision K
Optional Accessories
6
Fig. 45
1
2
3
4
5
using a
slotted screwdriver and then remove the cover
plate
in the
recess at the rear of the knife holder base.
• Connect the plug
of the
backlighting to the plug of the External power
supply unit Backlighting
.
Caution
When using super cassette clamp, the orientation is not set to position "0" when directional fixture for
specimen clamps is used with the rigid knife holder base, or the back-lighting illumination is attached.
The instrument may malfunction leading to delayed diagnosis.
• The orientation must be in position "0" and the cover for backlighting illumination must be detached.
• NEVER use the super cassette clamp with backlighting illumination.
• The force balance system has to be adjusted when using the super cassette clamp.
6.6
Top tray
Fig. 46
1
The top tray is mounted on the hood of the
on both
the top tray and the hood helps to orient the
installation of the top tray.
The top tray is for storage of the utensils used
during sectioning as well as the sectioned
specimens.
6.7
Universal microscope carrier
Tips
Unpack all accessories in the package and check for completeness.