HistoCore BIOCUT
Optional Accessories
6
6.1.2 Directional fixture for specimen clamps
Fig. 30
1
2
3
4
5
6
7
8
• Loosen the eccentric bolt
by turning
it counterclockwise.
• Completely unscrew the thrust piece
with a flat-tip screwdriver and pull it out with
.
• Completely unscrew the setscrews
and
.
• Install the directional fixture for specimen clamps
by tightening the screws in the bores
(2 screws are accessible through the bores)
using a Hexagon key No. 3. Insert the screws
in the bores as shown and screw
them in using a Hexagon key No. 3.
• Insert the spring
with the flatter side into the thrust piece
. Completely screw in the thrust
piece with a flat-tip screwdriver.
• Completely screw in the setscrews
6.1.3 Fine-directional fixture for specimen clamps
Fig. 31
1
3
4
2
• Before the fine-directional fixture for specimen
clamps can be mounted, loosen 4 screws
(Hexagon key No. 3) and carefully
remove the fixture for specimen clamps from the
the Hexagon key No. 3, fasten the baseplate to
the object head
• Now, screw the fine-directional fixture for
specimen clamps with the 4 screws
and the Hexagon key No. 3 onto the object
head.