HistoCore BIOCUT
Operation
5
Tip
The section thickness that has been set is not added to the selected trimming value.
If the section thickness that has been set is greater than the selected trimming value, the set section
thickness is fed.
5.1.5 Fine directional specimen holder fixture
Tip
In the quick clamping system of the directional specimen holder fixture, all specimen clamps
available as optional accessories can be used.
Fig. 15
1
2
3
4
4
The specimen orientation allows for simple position correction of the specimen surface when the
specimen is clamped into place.
The directional specimen holder fixture may be exchanged for a non-directional fixture (optional
accessory).
Display of the zero position
For better display of the zero position, the orientation has two red indicators
.
When both indicators are fully visible and both setscrews are in zero position at the same time (white
markings align with the arrows), the specimen is in zero position.