Version 1.7, Revision K
Setting up the instrument
4
Fig. 10
3
2
1
5
6
7
8
is loosely placed in the handwheel shaft
with a cable tie during transport.
1. Remove the cable tie
Caution
The feather key is lost when mounting the handwheel.
The instrument cannot be used, and it may cause the delay of diagnosis.
• Before assembling the handwheel, check the feather key and make sure that it is in the
handwheel shaft.
2. Place the handwheel
on the handwheel shaft
as shown.
3. Tighten the screw
located in the center hole of the handwheel with a Hexagon
4. Remove the cover foil from the self-adhesive cover disk
the handwheel.