HistoCore BIOCUT
Instrument Components and Specifications
3
General
Section thickness increments
• From 1 - 10 µm in 1 µm increments
• From 10 - 20 µm in 2 µm increments
• From 20 - 60 µm in 5 µm increments
Horizontal feed range
24 ± 2 mm
Vertical stroke length
70 ± 1 mm
Maximum cutting range without retraction
69 mm
Maximum sectioning area with retraction
62 mm
Maximum specimen block size for
large standard clamp (H x W x D)
Maximum specimen block size for
super cassette clamp (H x W x D)
55 x 50 x 30 mm
68 x 48 x 15 mm
Silent specimen retraction:
Approx. 40 µm. Can be turned off.
Trimming function with 2 settings
10 µm, 30 µm
Coarse feed wheel turn direction
clockwise or counter-clockwise
User selectable
Dimensions and weights
Width (including the handwheel and the
coarse feed wheel)
477 mm
Depth (including the section waste tray)
620 mm
Height (without the top tray)
295 mm
Height (with the top tray)
303 mm
Weight (without accessories)
Approx. 31 kg
Volume of the section waste tray
1400 ml
Optional accessories
Specimen orientation with zero postion
• Horizontal rotation: ± 8°
• Vertical rotation: ± 8°
Two-in-one blade holder
• Lateral displacement function
East-west movement
• Movement of blade holder base
• 3 positions
• North-south: ± 24 mm
Operating environment
Operating temperature range
+18 °C to +30 °C
Operating relative humidity
20 % to max. 80 % non-condensing