HistoCore BIOCUT
Safety
2
Warning
In the manual sectioning mode, the user cuts the specimen block by rotating the handwheel at a very
high speed.
It may lead to poor sectioning quality and even damage the specimen.
• In the manual sectioning mode, the rotation speed of the handwheel shall be no more than 60 rpm.
Warning
The rotation speed of the handwheel does not suit the hardness of the specimen.
Instrument damage and possibly specimen damage.
• The rotation speed of the handwheel must be adapted to suit the hardness of the specimen. For
harder specimens, use a slower speed.
Warning
The handwheel brake lever is not used properly, and the handwheel cannot be braked.
Serious injury of instrument/persons, or specimen damage.
• The handwheel brake lever must be exactly in the locked position. If the handwheel brake lever is
moved beyond this point, it is possible that the handwheel is no longer braked.
Warning
Fingers put between the specimen and the knife/blade after the handwheel is released when doing
fast manual trimming.
Operator can get cut and injured due to the turning of the handwheel when it is unlocked.
• Do not put fingers between the specimen and the knife/blade during trimming and sectioning.
Caution
After the knife/blade holder is laterally displaced, the specimen is not retracted and retrimmed.
Specimen damage.
• Retract the object head and retrim the specimen block each time after laterally displacing the
knife/blade holder.
Caution
Rotating the handwheel and the coarse feed wheel at the same time.
Specimen damage.
• Do not turn the coarse feed wheel whilst rotating the handwheel.
Caution
The feather key is lost when mounting the handwheel.
The instrument cannot be used, and it may cause the delay of diagnosis.
• Before assembling the handwheel, check the feather key and make sure that it is in the
handwheel shaft.