Version 1.7, Revision K
Safety
2
Note
Accessories and components experience corrosion due to using corrosive/strongly acidic/alkaline
reagent or solvent with the instrument or accessories, such as decalcified solution that contains
acid, ammonium hydroxide that contains alkali, etc.
Accessories can malfunction.
• Avoid corrosive/strongly acidic/alkaline reagent or solvent dripping onto the instrument surface or
accessories.
• If such reagent or solvent has dripped onto the instrument surface or accessories, wipe off the
residue and dry the accessories sufficiently as soon as possible.
• If such reagent or solvent is used frequently, carry out a thorough daily cleaning of the blade
holder, the universal cassette clamp (UCC), and other accessories if necessary.
2.2.4 Cleaning and maintenance
Warning
The liquids go into the interior of the instrument.
Serious injury of instrument/persons.
• Ensure that liquids do not enter the interior of the instrument during operation and maintenance.
Warning
Wiping the knife in the wrong direction during cleaning.
Serious injury of persons.
• Always wipe the knife from the back of the knife to the cutting edge.
Warning
During cleaning knife/blade holder, remove the parts from the drying chamber (65 °C).
Scalding hazard.
• Wear heat insulating gloves when removing the parts from the drying chamber (65 °C).
Caution
Use of improper solvents or cleaners, or sharp/hard tools to clean the instrument or accessories.
Potential instrument malfunction or delay of diagnosis.
• Do not use any solvents containing acetone or xylene for cleaning the instrument.
• When using cleaners, please comply with the safety instructions of the manufacturer and the
laboratory safety regulations.
• Never use a sharp or hard tool to scrape the instrument surface.
• Never soak the accessories in any cleaning solvent or water.
• Clean steel knives using an alcohol-based solution or acetone.
• For cleaning and removal of paraffin, do not use xylene or cleaning fluids containing alcohol (for
example glass cleaner).