8
–6
Triggers and When to Use Them
Hold-off by Time
This is the selection of
a minimum time for triggers (
Fig. 8–
3
). A trigger is generated when the trigger condition is met
after the selected delay from the last trigger. The timing for
the delay is initialized and started on each trigger. The hold-
off timeout should exceed the duration of the signal
displayed on screen. For instance, at a timebase setting of 1
ms/div, the time-out hold-off should at least exceed 10 ms.
Figure 8–3. Edge Trigger with Hold-off by Time.
Hold-off time
Hold-off time
Trigger Source: Positive Slope
Trigger can occur
Generated Trigger
Trigger
Event
Trigger
Event
Trigger
Event
Trigger initiates
hold-off timer
Trigger initiates
hold-off timer
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...