A–
28
Specifications
Post-trigger Delay:
0–10 000 divisions adjustable in 0.1 div
increments
Holdoff by Time:
10 ns–20 s
Holdoff by Events:
0–99 999 999 events
Internal Trigger Range:
±5 div
EXT Trigger Max Input:
Ø
50
Ω
±1 %:
±5 V DC (500 mW) or 5 V rms
Ø
1 M
Ω
/15 pF:
400 V max. (DC + peak AC
≤
10 kHz)
EXT Trigger Range:
±0.5 V (±5 V with Ext/10)
Trigger Timing:
Trigger Date and Time listed in “Memory
Status” menu
SMART Trigger Types
Signal or Pattern Width:
Triggers on width between two limits
of between
<
2.5 ns (1 ns typical) or pulse widths between
<
2.5 ns and 20 s exclusive
Signal or Pattern Interval:
Triggers on interval between two
limits of between 10 ns and 20 s
Dropout:
Triggers if the input signal drops out for a time-out
longer than 25 ns–20 s
State/Edge Qualified:
Triggers on any source only if a given
state or transition — number of events, time interval — on
another source
TV:
Selection of both line (up to 1500) and field number (up to
8) for PAL, SECAM, NTSC or nonstandard video
Exclusion Trigger:
Triggers only on shorter-than-normal
(defined) aberrations
Pattern:
Triggers on the logic combination of the five inputs CH
1, CH 2, CH 3, CH 4 and EXT Trigger, where each source can
be defined as High, Low or Don’t Know and the trigger as the
pattern’s beginning or end
Autosetup
AUTOSETUP button:
Sets timebase, trigger and sensitivity to
display wide range of repetitive signals — amplitude 2 mV–40 V;
frequency above 50 Hz; duty cycle greater than 0.1%
Autosetup Time:
Around two seconds
Vertical Find:
Automatically sets sensitivity and offset
Probes
Probe Model:
One PP005 probe supplied per channel (10:1,
10 M
Ω
//11 pF, 500 V max input); FET probes, purchased
separately, fully compatible with entire scope series
Probe calibration:
Max 1 V into 1 M
Ω
, 500 mV into 50
Ω
,
frequency and amplitude programmable, pulse or square wave
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...