D
–9
Parameter Measurement
P a r a m e t e r a n d w h a t i t d o e s
Definition
Notes
maximu
m
Measures highest point in waveform.
Unlike
top
, does
NOT
assume waveform
has two levels.
Highest value in
waveform
between cursors
(See Fig. D–1)
Gives similar result when
applied to time domain
waveform or histogram of
data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition. Computes
horizontal axis location of
rightmost non-zero bin of
histogram — not to be
confused with
maxp
.
mean
Average of
data
for time domain
waveform. Computed as centroid of
distribution for a histogram. But when
input is periodic time domain waveform,
computed on an integral number of
periods.
Average of
data
(See Fig. D–2)
Gives similar result when
applied to time domain
waveform or histogram of
data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition.
median
The average of base and top values.
Average of
base
and
top
(See Fig. D–2)
minimum
Measures the lowest point in a waveform.
Unlike
base
, does
NOT
assume waveform
has two levels.
Lowest value in
waveform
between cursors
(See Fig. D–1)
Gives similar result when
applied to time domain
waveform or histogram of data
of same waveform. But with
histograms, result may include
contributions from more than
one acquisition.
over
−
Overshoot negative: Amount of
overshoot following a falling edge, as
percentage of amplitude.
base
minimum
ampl
−
×
b
g
100
(See Fig. D–2)
Waveform must contain at
least one falling edge. On
signals
NOT
having two major
levels (triangle or saw-tooth
waves, for example), may
NOT
give predictable results.
over
+
Overshoot positive: Amount of overshoot
following a rising edge specified as
percentage of amplitude.
maximum
top
ampl
−
×
b
g
100
(See Fig. D–1)
Waveform must contain at
least one rising edge. On
signals
NOT
having two major
levels (triangle or saw-tooth
waves, for example), may
NOT
give predictable results.
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...