E–
5
ASCII Waveform Export
Plotting the data from a single-segment waveform requires the use
of a scatter plot based on the data in the first two columns with the
first column used as the X values.
The header created for the Spreadsheet format contains all the
information required to extract various elements of a sequence
waveform. The following Formulae may be used to extract
information such as the start and end row of the data for a given
segment, or the trigger time of a given segment.
SegmentStartRow
:= (
DesiredSegment
* D2) + B2 + 5
SegmentEndRow
:=
SegmentStartRow
+ D2 -1
TrigTime
= INDIRECT(ADDRESS(
DesiredSegment
+3;2;4))
TimeSinceFirstTrig
= INDIRECT(ADDRESS(
DesiredSegment
+3;3;4))
Plotting the data from all segments using a scatter plot will result in
all segments overlaid (similar to the scope’s display of sequence
traces in persistence mode).
Plotting a Single-
Segment Waveform
Extracting Segments
from Sequence
Waveforms
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...