8
–16
Triggers and When to Use Them
Interval Between Range
This Interval Trigger selects a maximum interval between the
two edges of the same slope (
Fig. 8–12
). The trigger is
generated on the second edge if it occurs within the selected
interval range. The timing for the interval is initialized and
restarted whenever the selected edge occurs. Intervals of
between 10 ns and 20 s can be selected.
Figure 8–12. Interval Trigger: Trigger if the interval is between the
selected ranges.
Interval width
Generated Trigger
Interval
width
T2
Selected
ranges
Trigger can occur
T1
T2
T1
Trigger Source: Positive Slope
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...