8
–4
Triggers and When to Use Them
Trigger Source
The trigger source may be:
Ø
The acquisition channel signal (CH 1, CH 2 or CH 3, CH 4
on four-channel models) conditioned for the overall voltage
gain, coupling, and bandwidth.
Ø
The line voltage that powers the oscilloscope (LINE). This
can be used to provide a stable display of signals
synchronous with the power line. Coupling and level are not
relevant for this selection.
Ø
The signal applied to the EXT BNC connector (EXT). This
can be used to trigger the oscilloscope within a range of
±
0.5 V, or
±
5 V with EXT/10 as trigger source.
Level
Level
defines the source voltage at which the trigger circuit will
generate an event (a change in the input signal that satisfies the
trigger conditions). The selected trigger level is associated with
the chosen trigger source. Note that the trigger level is specified
in volts and is normally unchanged when the vertical gain or
offset is modified.
The Amplitude and Range of the trigger level are limited as
follows:
Ø
±
5 screen divisions with a channel as trigger source
Ø
±
0.5 V with EXT as trigger source
Ø
±
5 V with EXT/10 as trigger source
Ø
None with LINE as trigger source (zero crossing is used).
Coupling
This is the particular type of signal coupling at the input of the
trigger circuit. As with the trigger level, the coupling can be
independently selected for each source. Thus changing the
trigger source can change the coupling. The types of coupling
able to be selected are:
Note:
Once specified, Trigger Level and Coupling are
the only parameters that pass unchanged from mode
to mode for each trigger source.
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...