Timebase Modes and Setup
Timebase Sampling Modes
Depending on the timebase, any of three sampling modes
can be chosen: Single-Shot, Random Interleaved Sampling
(RIS) or Roll Mode. Furthermore, for timebases suitable for
either Single-Shot or Roll Mode, the acquisition memory
can be subdivided into user-defined segments to give
Sequence Mode. Channels can also be combined to boost
sample rate and record length.
Single-Shot
Single-Shot is the digital oscilloscope’s basic acquisition
technique and other timebase modes make use it.
An acquired waveform consists of a series of measured voltage
values sampled at a uniform rate on the input signal. The
acquisition, a single series of measured data values associated
with one trigger event, is typically stopped at a fixed time after
the arrival of the event, this being determined by the trigger
delay. The time of the trigger event is measured using the
timebase clock. The horizontal position of a waveform is
determined using the trigger event as the definition of time zero.
Waveform display is also carried out using this definition.
Because each channel has its own ADC, the voltage on each
input channel is sampled and measured at the same instant.
This allows very reliable time measurements between different
channels.
Trigger delay can be selected anywhere within a range that
allows the waveform to be sampled from well before the trigger
event up to the moment it occurs (100 % pre-trigger), or at the
equivalent of 10 000 divisions (at the current time/div) after the
trigger.
For fast timebase settings the ADCs’ maximum single-shot
sampling rate is used (on one and each channel, with higher
sampling rates achieved by combining channels —
see page
7–4
). For slower timebases, the sampling rate is decreased and
the number of data samples maintained. (
See
Appendix A
for
details
).
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...