7
–7
TIMEBASE EXTERNAL
— appears when an External clock mode is chosen.
Sampling
This menu is inactive when the external sample clock is being used.
Only single-shot acquisition is available (
see below
).
Sample Clock
For selecting a description of the signal applied to the EXT BNC
connector for the sample clock up to 100 MHz. The rising edge of the
signal is used to clock the ADCs of the oscilloscope. The effective
thresholds for sampling the input are
*
:
ECL.....................................
−
1.3 V
0V ....................................... 0.0 V
TTL...................................... +1.5 V
(With CKTRIG Option ONLY)
RP (Rear Panel) specifies that
the 50–500 MHz external clock connected to the rear panel be
used as the sample clock (
see
CKTRIG Manual
for details
).
External
To select the input coupling for the external clock signal.
Sequence
Offers Sequence Mode. The corresponding knob is used to adjust the
number of segments. Neither the trigger time stamps nor the AUTO
sequence time-out feature are available when the external clock is in
use. Nor is the inter-segment dead time guaranteed.
Record
To select the desired number of samples for the single-shot
acquisition.
See Appendix A for model maximums.
*
External clock modes are available only if the EXT trigger is not
the trigger source.
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...