![LeCroy 9300C Series Operator'S Manual Download Page 230](http://html1.mh-extra.com/html/lecroy/9300c-series/9300c-series_operators-manual_1866306230.webp)
C
–17
FFT Glossary
Glossary
Defines the terms frequently used in FFT spectrum analysis
and relates them to the oscilloscope.
Aliasing
If the input signal to a sampling acquisition system contains
components whose frequency is greater than the Nyquist
frequency (half the sampling frequency), there will be less than
two samples per signal period. The result is that the contribution
of these components to the sampled waveform is
indistinguishable from that of components below the Nyquist
frequency. This is
aliasing
.
The timebase and transform-size should be selected so that the
resulting Nyquist frequency is higher than the highest significant
component in the time-domain record.
Coherent Gain
The normalized coherent gain of a filter corresponding to each
window function is 1.0 (0 dB) for a rectangular window and less
than 1.0 for other windows. It defines the loss of signal energy
due to the multiplication by the window function. This loss is
compensated in the oscilloscope. This table lists the values for
the implemented windows.
Window Frequency-Domain Parameters
Window Type
Highest Side
Lobe
(dB)
Scallop Loss
(dB)
ENBW
(bins)
Coherent Gain
(dB)
Rectangular
–13
3.92
1.0
0.0
von Hann
–32
1.42
1.5
– 6.02
Hamming
–43
1.78
1.37
–5.35
Flat Top
–44
0.01
2.96
–11.05
Blackman–Harris
–67
1.13
1.71
–7.53
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...