CURSORS/MEASURE & Parameters
CHANGE TEST — Param (Changing Limits for Pass/Fail Tests on Parameters)
On line
To select for modification up to five different parameters, “
1
,
2
,
3
,
4
or
5
”.
Test on
To Select “
Param
” or “
---
(No Test)
” if no test required on the
selected line (
“Mask” selection, page 14–17
).
choose
For setting to “
Limit
” (
“Param” 14–15
).
DELETE ALL TESTS
For deleting all tests previously selected.
True if
For selecting the adequate relation — smaller than “
<
” , or greater
than “
>
”.
limit
To make the choice of one of three modifications of a limit: its
mantissa, exponent, and the number of digits for the representation
of its mantissa. The corresponding menu button is used to select,
and the associated knob to modify the number in that field.
SET TO LATEST VALUE
To set the limit to the latest measured value — a starting value for
the final adjustment.
Summary of Contents for 9300C Series
Page 1: ...Operator s Manual LeCroy 9300C Series Digital Oscilloscopes Revision A January 1998 ...
Page 8: ...1 3 billed for parts and labor related to the repair work and for shipping ...
Page 18: ...4 1 4 Introduction to the Controls Two Channel Front Panel ...
Page 19: ...4 2 Introduction to the Controls Four Channel Front Panel ...
Page 92: ......
Page 111: ...11 3 Dual Quad ...
Page 112: ...11 4 Display Parameter s XY only ...
Page 113: ...11 5 XY Single XY Dual ...
Page 138: ...12 21 Ø FLASH UPDATE Offering the Update from and Update Program menus illustrated next page ...
Page 139: ...12 22 UTILITIES The full screen warning message shown when FLASH UPDATE has been selected ...
Page 142: ...13 2 WAVEFORM STORE RECALL ...
Page 182: ......
Page 189: ...A 15 9344C Series 9350C Series 9354C Series 9354CTM 50 000x 9344CL 9350CL 9354CL 100 000x ...
Page 252: ...E 6 Appendix E ...
Page 275: ...Index 9370C 9374C Series A 18 9384C Series A 24 Zoom menu 10 7 Zoom of Math Functions 10 1 ...