14. STATUS DISPLAY
179
14.8.8
Setting the Luminance Error Levels
To set the luminance error detection levels and the detection area, follow the procedures
below.
These settings are available when F•1 LUMA ERROR is set to % or mV.
The settings that you make here also apply to the Y bar in the 5-bar display.
●
Setting the Upper Error Detection Level
An error is detected when the input signal level exceeds the specified value.
If you press F•D 1, the setting will be set to its default value of 109.4 % or 765.8 mV.
Procedure
STATUS
→
F•4 ERROR CONFIG
→
F•3 GAMUT ERROR
→
F•4 LUMA ERROR
→
F•2 UPPER LEVEL
→
F•D 1 VARIABLE:
90.8% - 109.4% (When LUMA ERROR is set to %)
635.6mV - 765.8mV (When LUMA ERROR is set to mV)
●
Setting the Lower Error Detection Level
An error is detected when the input signal level goes below the specified value.
If you press F•D 1, the setting will be set to its default value of -7.2 % or -50.4 mV.
Procedure
STATUS
→
F•4 ERROR CONFIG
→
F•3 GAMUT ERROR
→
F•4 LUMA ERROR
→
F•3 LOWER LEVEL
→
F•D 1 VARIABLE:
-7.2% - 6.1% (When LUMA ERROR is set to %)
-50.4mV - 42.7mV (When LUMA ERROR is set to mV)
●
Setting the Error Detection Area
An error is detected when an error greater than or equal to the specified value occurs in the
active picture. If you press F•D 1, the setting will be set to its default value of 1.0 %.
Procedure
STATUS
→
F•4 ERROR CONFIG
→
F•3 GAMUT ERROR
→
F•4 LUMA ERROR
→
F•4 AREA
→
F•D 1 VARIABLE: 0.0% - 1.0% - 5.0%
14.9 Clearing
Errors
To reset the error counts on the status display and on the error counters to zero, follow the
procedure below.
Also, the LAPSED indication at the upper left of the screen is reset to 00:00:00.
Procedure
STATUS
→
F•5 ERROR CLEAR