Maintenance
28.04.10/ YACE0088
9
Maintenance
9.1
Alarms, warnings and errors
Your ECO thermostat triggers alarms, warnings or error messages as necessary.
All warnings, alarms or error messages are shown in plain text in the display.
Alarms
:
Alarms are relevant to safety. Pump, heating and chiller switch off.
Warnings
:
Warnings are normally not relevant to safety. The device continues to run.
Errors
:
If a malfunction occurs, switch off the unit at the mains switch. If the malfunction recurs
after switching on the device, contact LAUDA Service Constant Temperature Equip-
ment (
Ì
9.6) or your local service organisation.
Once the cause has been rectified, alarms and warnings can be cleared with
.
Warnings can be ignored with
without the message periodically appearing again.
9.1.1
Overtemperature protection alarm and overtemperature protection check
The devices are rated for use with flammable and non-flammable liquids according to
DIN EN 61010-2-010.
−
Set the overtemperature switch-off point as described in (
Ì
7.4.1). Recommended setting: 5 K above the desired maxi-
mum bath temperature.
−
Important!!
The overtemperature switch-off point T
max
is con-
trolled by a system which operates independently of the bath
control.
−
The overtemperature switch-off point is displayed on pressing
in the display.
−
Do not set the switch-off point higher than 25 K below the fire point of the heat transfer