![Lauda RE 1050 G Operating Instructions Manual Download Page 55](http://html1.mh-extra.com/html/lauda/re-1050-g/re-1050-g_operating-instructions-manual_4088318055.webp)
ECO GOLD
28.04.10/ YACE0088
Putting the device into operation
55
−
The programmer edit window contains the following parameters:
No.
:
Program segment number
Tend
:
Final temperature to be attained
hh
:
Time in hours (hh) in which the specified temperature is to be attained
mm
:
Time in minutes (mm) in which the specified temperature is to be attained
−
If the value "0" is entered in the fields "hh" and "mm", the set value is accepted immediately and the bath
temperature approached as quickly as possible.
Tolerance
:
Defines how exactly the final temperature is to be attained before the next segment it proc-
essed.
−
If the tolerance range is selected too small in the "Tol." field, the program might not continue, because
the required tolerance is not achieved.
Pump
:
Pump level at which the segment is to be processed.
S1, S2, S3
:
Switching contacts of the contact module (if present) can be programmed here. Contact
modules are available as accessories (
Ì
10). The setting "- - -" stands for no change to the preceding seg-
ment, i.e. if "- - -" is present in all fields, the contact setting of the start setup or that before the program start
is retained.
Inserting a new segment
−
With or
go to the segment number after which the new
segment is to be inserted.
−
The new segment is inserted on pressing
(new). You can
edit it as described above.
Deleting a segment
−
With or
go to the segment number which is to be de-
leted.
−
The new segment is deleted on pressing
(delete).
No. Tend
hh
mm
Tolerance
Start
30.00 -- --
0.1
2 50.00 0 20
0.0
3
50.00 0
20
0.0
4 70.00 0 20
0.1
5 60.00 0 30
0.0
6 30.00 0 0
0.0
ESC
סּ
new
delete
No. Tend
hh
mm
Tolerance
Start
30.00 -- --
0.1
2 50.00 0 20
0.0
3
50.00 0
20
0.0
4 70.00 0 20
0.1
5
60.00 0
30
0.0
ESC
סּ
new
delete