Putting the device into operation
59
−
Access to the menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Control
Æ
Control parameter
Æ
intern Pt100
. The adjacent menu
window appears. Apart from the control parameters the
currently set values are displayed.
−
Under the menu point "Tv manual/auto" you can select
between manual and automatic entry using
. The se-
lection "automatic" is displayed in the menu line by (auto).
−
If "automatic" is selected, the entry is blocked for the pa-
rameters Tv and Td.
−
Select the parameters with
or
and confirm with
.
−
The respective edit window appears with Min and Max
figures for the parameter values Xp, Tn, Tv and Td.
−
By
pressing or
you can also select numbers individu-
ally and change them with
or
.
−
You confirm the set value by pressing
.
−
You quit the window without changes with
(ESC).
ESC
סּ
OK
Standby
intern Pt100
Xp
10.0
Tn
30
Tv manual/auto auto
Tv
24(auto)
Td
4(auto)
ESC
סּ
OK
- - -
Xp
Max:
30,0
Min:
0,1
10,0