Putting the device into operation
53
Edited programming example (refer to dashed curve in graph 1)
No. Tend hh mm Tolerance
No. Pump S1 S2 S3
Start 30.00 --
--
0.1
Start
2
off off off
2 50.00 0 20
0.0
2
2
off off off
3
50.00
0 20
0.1
e
3 3
off
off
off
4 70.00 0 20
0.0 4
4
off
off
off
5 70.00 0 10 0.8
e
4 2
off
off
off
6 60.00 0 30
0.0
5
2
off off off
7 30.00 0 0
0.0
6 2
off
off
off
Insert new segment (
Ì
e
Change segment time or tolerance (
Ì
The field "Tolerance"
(Refer to programming table above and graph below):
−
It facilitates exact conformance to the dwell time at a specified temperature. Segment 1 is
not processed until the actual temperature is within the tolerance band
n
, so that the ramp
(Segment 2) starts delayed at
o
.
−
A tolerance range which is too tight can however also cause undesired delays.
In particular
with external control
the range should not be chosen too tightly. In Segment 5 a larger tol-
erance has been entered, so that the desired time of ten minutes is maintained even with
settling action
p
.
−
Only flat (slow) ramps should be programmed where necessary with a tolerance range.
Steep ramps which lie close to the maximum possible heating or cooling rates of the ther-
mostat may be severely delayed by a tolerance range that is too tight (here in Segment 2)
q
.
Example of the effect of the tolerance input with
external bath control:
The setpoint temperature of the programmer is
shown in grey.
The actual temperature in the external bath container is
represented as a continuous line.
7.8.2
Creating or changing a program (editing)
Here, there are the following functions:
•
Create a program.
•
Display the program data of a saved program and change the segment data.
•
Insert or append a new segment.
•
Delete
a
segment.
°C
80
70
60
50
40
30
20
10
0 20 40 60 80
100
120
t/min