Preparations
28.04.10/ YACE0088
6.4
Heat transfer liquids and hoses
Heat transfer liquids – approved by LAUDA
LAUDA
designation
Operating
temperature
range
Chemical
designation
Vis-
cosity
(kin)
Viscosity
(kin) at
temperature
Fire
point
Container size
Catalogue number
°C
mm²/s
(20 °C)
mm²/s
°C
5 L
10 L
20 L
Kryo 51
-50…120
Silicone oil
5
34 at -50 °C
> 160
LZB 121 LZB 221 LZB 321
Kryo 30
-30…90
Monoethylene glycol
/water
4
50 at -25 °C
--
LZB 109 LZB 209 LZB 309
Kryo 20
-20…180
Silicone oil
11
28 at -20 °C
> 230
LZB 116 LZB 216 LZB 316
Therm 180
0…180
Silicone oil
23
36 at 0 °C
> 288
LZB 115 LTB 214 LZB 314
Aqua 90
5...90 Decalcified
water
1
--
--
LZB 120 LZB 220 LZB 320
Ultra 350
30...200
Synth. heat carrier
47
28 at 30 °C
> 240
LZB 107 LZB 207 LZB 307
Therm 240
50…240
Silicone oil
125
45 at 50 °C
> 378
LZB 122 LZB 222 LZB 322
Therm 200
60...200
Silicone oil
5
4
28 at 60 °C
>
362
LZB 117 LZB 217
LZ
B 317
{
At higher temperatures vaporisation losses occur. In this case use a bath cover (
Ì
distilled water or pure demineralised water only after adding 0.1 g of soda (Na
2
CO
3
sodium
carbonate)/litre of water. Otherwise there is the risk of corrosion!
The proportion of water reduces with longer working at high temperatures and the mixture be-
comes flammable (flash point 128 °C). Check the mixing ratio using a hydrometer.
Do not use in conjunction with EPDM hose.
−
When choosing the heat transfer liquid, it must be noted that at the lower limit of the operating
temperature range a worsening of the tempering properties is to be expected due to the in-
creasing viscosity. Therefore, only use the full operating temperature ranges where neces-
sary.
−
The working ranges of the heat carrier liquids and hoses are general figures which can be
tightened due to the operating temperature range of the devices.
−
Do not use any contaminated heat transfer liquids. Contamination of the pump chamber may
lead to the pump jamming and the device then switching off.
−
Never use silicone oil with silicone hoses.
Safety data sheets can be ordered on request.