Putting the device into operation
28.04.10/ YACE0088
7.5.9
Limiting the mains current consumption
If your mains fusing is below 16 A, the current consumption can be reduced in steps from 16 A to 8 A.
The maximum heating power is then also reduced correspondingly. Here, take into consideration
whether other loads are connected to the same fused circuit or whether your ECO thermostat is the only
load.
−
Access to the main menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Basic setup
.
The adjacent menu window appears.
−
As well as
Curr.Consumpt.
the set value is displayed.
−
Select
Curr.Consumpt.
with
.
−
Change the value with
or
.
−
By
pressing or
you can also select numbers indi-
vidually and change them with
or
.
−
You confirm the set value by pressing
.
−
You quit the window without changes with
(ESC).
ESC
סּ
OK
Standby
Basic setup
Acoustic
signal
►
Display
►
Autostart on
►
Resolution
Curr.Consumpt. 16.0A
DLK
connected
Language
►
ESC
סּ
OK
Standby
Curr.Consumpt.
Max:16,0
Min:8,0
16,00