Putting the device into operation
45
7.5.10
Entering the offset of the displayed temperature (calibration)
If a deviation is found when checking with a calibrated reference thermometer, e.g. from the LAUDA
DigiCal series, the offset of the internal measuring chain can be adjusted with the following function.
−
Access to the main menu level is obtained by pressing
−
Selection and confirmation of
Æ
Setup
Æ
Calibration
.
The adjacent menu window appears.
−
Select "Calibration" with
or
and confirm with
. The entry window appears.
−
Change the value with
or
.
−
By
pressing or
you can also select numbers indi-
vidually and change them with
or
.
−
By
pressing
(+/-) the arithmetic sign "+" or "-" can be
selected.
−
You confirm the set value by pressing
.
−
By
pressing
(ESC) you are returned to the menu level
without the actual value being changed.
ESC
סּ
OK
Standby
Calibration
Calibration
.
Factory calibration
►
ESC
סּ
OK
+/-
Temp. of Reference
Max:
28,00
Min:
22,00
25,00