![Lauda RE 1050 G Operating Instructions Manual Download Page 52](http://html1.mh-extra.com/html/lauda/re-1050-g/re-1050-g_operating-instructions-manual_4088318052.webp)
ECO GOLD
52
Putting the device into operation
28.04.10/ YACE0088
7.8
Programmer
The programming function enables you to save five temperature/time programs. The programs consist of
several temperature/time segments. This also includes how often the program is to be run (loops). The
sum of all segments of all programs can be up to 150. Step changes in temperature, i.e. the time is zero,
or also temperature holding phases, i.e. the temperature at the start and end of a segment is the same,
are possible. On starting the current set value is taken as the starting value of the first segment.
If the pump level is to be changed in the course of the program, then the specified pump level is entered
in the respective program line. If the pump level is to remain unchanged, "0" is entered and it is indicated
as "---" in the display.
7.8.1
Programming example
No. Tend hh
mm
Tolerance
No. Pump S1 S2 S3
Start 30.00 -- --
0.1
Start
2
off off off
2 50.00 0 20 0.0
2
2 off off off
3 70.00 0 40 0.0
3
3 off off off
4 70.00 0 10 0.1
4
4 off off off
5 60.00 0 30 0.0
5
2 off off off
6 30.00 0 0
0.0
6 2 off
off
off
No time specification is possible in segment 1 (starting segment). The temperature of the first
segment is attained as quickly as possible in order to switch to segment 2 after reaching the
set tolerance.
Cooling time dependent on de-
vice type, load etc.
Graph 1