Putting the device into operation
51
7.7.2
Setpoint offset operating mode (Diff.set/actual)
It is possible to apply an offset value to the temperature, which is provided by an external temperature
sensor and to process it as the set value. The bath temperature can therefore be operated, for example,
-15 °C below the temperature of a reactor measured by the external temperature sensor.
−
Access to the menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Control
Æ
Setpoint offset
. Adjacent menu window appears.
−
As well as
Diff.set/actual value
the currently set offset value
is displayed.
−
The adjacent menu window appears on selecting the menu
point "Offset source".
−
Select the set-value source with
or
and confirm it
with
.
−
The setpoint offset is deactivated with "off".
−
Select
Diff.set/actual value
with
or
and confirm
with
. The adjacent menu window appears.
−
The minimum and maximum possible offset values and the
current offset value are displayed.
−
Change the value with
or
.
−
By
pressing or
you can also select numbers indi-
vidually and change them with
or
.
−
By
pressing
(+/-) the arithmetic sign "+" or "-" can be
selected.
−
You confirm the set value by pressing
.
−
You quit the window without changes with
(ESC).
ESC
סּ
OK
Standby
Setpoint offset
Offset source
►
Diff.set/actual value 0.0
ESC
סּ
OK
+/-
Offset value
Diff.set/actual value
Max:500,0
Min:-500,0
0,0
ESC
סּ
OK
Standby
Offset source
off
extern Pt100