Putting the device into operation
39
7.5
Other settings
7.5.1
Defining temperature limits
With this function it is possible to define a minimum and maximum temperature within which the controller
operates. If, for example, you are using water as the heat transfer liquid, +95 °C would be practicable as
the maximum temperature and +5 °C the minimum temperature.
−
Access to the main menu level is obtained by pressing
.
−
The adjacent menu window appears by selecting and
confirming
Æ
Setup
Æ
Limits
.
−
Select the lower (Til) or upper (Tih) limit with
or
and confirm it with
. In the entry window the cursor
flashes below the value to be changed. The permissible
adjustment range is indicated with Min and Max.
−
Change the value with
or
.
−
By
pressing or
you can also select numbers indi-
vidually and change them with
or
.
−
By
pressing
(+/-) the arithmetic sign "+" or "-" can be
selected.
−
You confirm the set value by pressing
.
−
By
pressing
(ESC) you are returned to the menu
level without the set-point being changed.
ESC
סּ
OK
Standby
Limits
T il -30.0°C
T il 202.0°C
ESC
סּ
OK
Standby
T il
Lower limit
Max:202.0
Min:-30.0
-30,0