34
LABORIE Urocap™ IV Owner’s Manual UC4-UM01
The comparison is provided in a percent difference from the table value. Following is the mathematical formula
used in the calculations:
Peak Flow:
(
)
(
)
%
100
,
,
∗
−
Sex
Volume
n
PeakFlowMi
Sex
Volume
n
PeakFlowMi
PeakFlow
Mean Flow:
(
)
(
)
%
100
,
,
∗
−
Sex
Volume
n
MeanFlowMi
Sex
Volume
n
MeanFlowMi
MeanFlow
Flow Time:
(
)
(
)
%
100
,
,
∗
−
Sex
Volume
uritionMax
TimeOfMict
FlowTime
Sex
Volume
uritionMax
TimeOfMict
Time to Peak Flow:
(
)
(
)
%
100
,
,
∗
−
Sex
Volume
xFlowMax
TimeOfToMa
Flow
TimeToPeak
Sex
Volume
xFlowMax
TimeOfToMa
These percent values are found in the patient statistics of the Voiding Summary printout. The M% represents the
male values and the F% represents the female values.
The comparison with normal parameters should not be used for diagnostic purposes. It is
recommended to repeat the study if the total voided volume is less than 55 ml; normal parameters
are not available below this volume. Statistic values cannot be given if the voided volume is less than
55 cc and more than 555 cc.
Parameters of urinary flow were
analyzed from 150 curves from subjects between 20 and 40 years old.
The minimum normal for a given volume appears to
be the most practical value with which to
compare patient flow rate.
6.4.5
UROFLOW OPTIONS
From the
Uroflow Options box you can:
•
Customize how the summary is calculated.
•
Make changes to the channels and configuration parameters.
TO CHANGE PARAMETERS:
1.
Click Options located in the Voiding Summary box.
2.
Refer to the following list to change any parameters in the Uroflow Options dialog box.
PARAMETER
DESCRIPTION
Default Channels
Flow and Volume
Range
Default range is from start to end of test
Threshold
The minimum flow rate value used in calculating the Voiding Summary.
Anything below this threshold is considered noise or insignificant flow.
Enable Artifact Recognition
Select check box to enable
Artifact Recognition Threshold
Indicates the maximum acceptable change in Flow (acceleration) that is
physiologically possible.
OEM Uroflow Transducer
Original Equipment Manufacturer (for optional non-Laborie Uroflow
transducers).