LABORIE Urocap™ IV Owner’s Manual UC4-UM01
Uroflowmeter signal
does not respond
Urocap™ IV improperly connected
Connect the Urocap™ IV to the software
Uroflowmeter signal
shows vibration and/or
spike patterns
Plastic beaker is touching funnel
Reposition the Uroflowmeter and check
again.
Patient has touched uroflowmeter
with their feet
Ask patient to remain calm during procedure
Floor is unsteady or uneven
Move to a more solid foundation
Incorrect flow or
volume readings
Beaker not properly seated on dish
Adjust beaker position
Funnel is touching the beaker
Adjust commode chair or Urocap™ IV
Incorrect beaker is in use
Use beakers supplied by LABORIE only
Error message on
screen reads:
“Calibration
Failure…(Flow out of
range)”
Water was poured into beaker too
quickly
Pour water slowly and at a steady rate
More than 500 ml poured into beaker 500 ml must be poured into the beaker
Error message on
screen reads:
“Calibration
Failure…(Volume delta
out of range)”
Less than 500 ml of water poured
into the beaker
500 ml must be poured into the beaker
Table 17 – Troubleshooting Urocap™ IV Usage
11.2
BLUETOOTH CONNECTION
SYMPTOM(S)
POSSIBLE CAUSE(S)
CHECK/ CORRECTIVE ACTION(S)
Unable to connect via
Bluetooth
Broken connection
Press the system reset button on the bottom
of the device
Ensure Bluetooth is on by clicking the device
icon in the Urocap™ IV Device Manager
Reduce the distance between the device and
the computer (maximum distance is 10m/33’)
Remove physical barriers such as walls or
people
Table 18 – Troubleshooting Bluetooth Connectivity