52
LABORIE Urocap™ IV Owner’s Manual UC4-UM01
12
APPENDICES
APPENDIX A.
SPECIFICATIONS AND SYSTEM INFORMATION
Upon request, LABORIE will make available any circuit diagrams, component part lists, and other technical
documentation directly related to the Urocap™ IV.
Urocap™ IV (TRA851)
Dimensions: 6.3”(16.0cm) L X 6.3”(16.0cm) W X 2.6”(6.6cm) H
Weight: 0.7lbs (0.3kg)
Flow Range: 0 to 50mL/s; Volume Range: 0 to 1500mL
Output Channels: Flow; Volume
Sampling Rate: Flow=20Hz; Volume=20Hz
Charger: Connect to SL Power Electronics Corp. Model: MW172KB0540F02 or
ME20A0540F02 PSU only.
Power Supply: POW044 - 5V, 3A
Transportation and
Storage Conditions
Temperature: -29ºC to +60ºC
Humidity: uncontrolled to 85% relative humidity
Operating Conditions
Temperature: +10ºC to +40ºC (50ºF to 104ºF)
Humidity: 30% to 75% relative humidity (non-condensing)
Pressure: 700 hPa to 1060 hPa (up to 2000m)
Table 21 – Equipment Specifications
CLASSIFICATIONS
Table 22 – Classification
IEC 60601-1:
Class I Equipment Type BF Applied Parts
Mode of Operation:
Continuous; Equipment not suitable for use in the presence of a Flammable,
Anaesthetic Mixture, with Air, or Oxygen, or Nitrous Oxide.
Degree of Protection:
The Urocap™ IV uroflowmeter enclosure is classified IP54 according to degree of
protection against ingress of water and particulate matter as per the test requirements
of IEC 60529. With this IP (
International Protection
) rating, it means that the Urocap™
IV uroflowmeter enclosure:
Protects users using tools 1.0 mm or larger from accessing hazardous parts, and
protects equipment from ingress of dust (signified by the rating code 5).
Protects equipment from the harmful effects of water splashing from any direction
(signified by the rating code 4).
NOTE: If applicable to your uroflowmeter model, the IP rating will be visible on the
device label.