48
LABORIE Urocap™ IV Owner’s Manual UC4-UM01
9
CALIBRATION
STEP
1
Plug the Urocap™ IV into the charger and plug
the charger into an outlet.
2
Start the UDS120 Goby software and double-click
the Urocap™ IV icon in the main window to open
the
Urocap™ IV
Device Manager
window.
3
Click the Setup button in the
Urocap
section.
4
In the
Urocap Setup
window click the Device
Calibration button.
5
Place an empty beaker on the Urocap™ IV and
then click Apply.
6
Follow the on-screen instructions to complete the
calibration.
7
Click Close when calibration complete.
Figure 21 – Calibrating the Urocap™ IV with UDS120
10
QUICK REFERENCE AND FAQ
How do I check that the Urocap™ IV is connected?
You can check equipment connection one of two ways:
i.
The equipment’s LED lights alert you of its status.
ii.
Open the Device Manager and see if connections are enabled.
How do I know when I need to charge the Urocap™ IV?
The LED lights on the
Urocap™ IV
, as well as the Device Manager window, will alert you of the device’s battery
status. See Section 5.1 for a chart listing LED light status.
How do I charge the Urocap™ IV?
See Section 3.1 for more information about charging the devices.
Can I still use the Urocap™ IV while battery is charging?