Manual Operation
4-11
4.3.1 Add new protocol
1.
From the
P
ROTOCOLS
window, click the
A
DD
button. The
New Protocol
wizard will begin.
2.
Enter the protocol name (for example, 384LDV Aqueous P 10 nL) and
select the source and destination microplate types from the drop-down
lists.
Figure 4.7 Entering protocol name
3.
Click the
N
EXT
button. The second window will open.
Figure 4.8 Add new protocol wizard, page 2
4.
Enter the number of copies (for example, 5). Each copy represents a
single plate transfer.
5.
Enter the liquid transfer volume (for example, 10 nL). If you enter the
transfer volume directly instead of using the increment/decrement
buttons, remember that the value must be an even multiple of the
minimum transfer volume (2.5 nL). If you enter any other value, you will
see an error message.