Maintenance and Service
6-15
6.2.5 Replace the water filter
During plate survey and fluid transfer, the coupler fluid is exposed to dust
and particles from the air or from the bottom of the source microplates.
The inline water filter on the chiller tubing filters out dust and particles as
the coupling fluid returns to the chiller.
To ensure that the coupler fluid is filtered effectively, Labcyte will replace
the water filter every six months. If the water filter appears discolored or
cracked before the six month maintenance is due, replace it immediately.
Figure 6.4 Water filter in good condition versus discolored
Tools and materials needed
Inline water filter (P/N 001-4561)
Nylon hose clamps 11/32 in, if needed (P/N 001-4045)
Compressed air or vacuum with brush
To replace the water filter:
1.
Turn off the coupling fluid pump from the Diagnostics window.
2.
Remove the hose clamps from the ends of the water filter.
3.
Gently pull the water filter off the tubing.
4.
Insert the new water filter.
Note:
Ensure that the arrow on the water filter is pointing in the
direction of the water flow (away from the instrument).
Figure 6.5 Water filter direction.
5.
Secure the new water filter with the hose clamps.
6.
Turn on the coupling fluid pump.
Summary of Contents for Echo 520
Page 1: ...For Echo models 520 550 555 ...
Page 20: ...xx Echo Liquid Handler User Manual ...
Page 58: ...2 14 Echo Liquid Handler User Manual ...
Page 82: ...3 24 Echo Liquid Handler User Manual ...
Page 100: ...4 18 Echo Liquid Handler User Manual ...
Page 160: ...5 60 Echo Liquid Handler User Manual ...
Page 192: ...A 8 Echo Liquid Handler User Manual ...
Page 199: ......