5-46
Echo Liquid Handler User Manual
If register calibration is required, use the following procedure:
1.
Click
R
EGISTER
in the
Calibration
window. The
Plate Registration
window opens.
Figure 5.42 Plate registration window
2.
Select the source plate type from the drop-down menu.
3.
Click the
P
RESENT
button to extend the source plate gripper stage.
4.
Place the source plate on the plate gripper.
5.
Click the
R
ETRACT
button to draw the plate into the instrument.
6.
Click the
R
EGISTER
button. The Echo liquid handler calibrates the well
position and displays the new well center values.
7.
Click the
A
PPLY
L
ABWARE
button at the bottom of the dialog box, then
Y
ES
in the
Confirmation Update
dialog box, to update the database.
Figure 5.43 Plate registration confirmed