73
Service Components
– Humidifier Assembly (Item 16)
Always drain water from the Humidifier before service. Instructions for draining the
Humidifier can be found in
Moving the Chamber
earlier in this section. Serviceable
parts are located in the Humidifier Compartment, these components can be replaced as
follows:
ALWAYS UNPLUG THE POWER CORD BEFORE SERVICING THIS PRODUCT!
27. Float Switch
– P/N 3192601
Disconnect wires, open Humidifier Lid, and unscrew nut holding Float Switch in place.
28. Humidifier Valve Assy
– P/N 3185001P
Remove Humidifier Wire Connector, then remove Humidifier by removing the two (2)
Humidifier Basin Screws shown in Fig. 7-18. Disconnect 2-pin wire harness connector
to Valve Assembly, then remove the three (3) nuts holding the Valve Assembly to the
wall. Tilt top of Valve Assembly away from the wall and lift out.
27
28
29
30
HUMIDIFIER
BASIN
SCREWS (2)
HUMIDIFIER
WIRE
CONNECTOR
Figure 7-18