33
Chamber Configuration
Before using the chamber to process evidence, it is recommended that you configure
the
chamber’s settings to your preference.
Keypad button presses are shown as
[BLUE WITH BRACKETS]
. Menu screen
selections are shown as
green italics
.
On the Main Menu screen, Select
SETTINGS
by moving the red arrow down to Settings
menu option, and press
[OK]
. The Settings Menu (see Fig. 6-5) displays seven sub-
menus, several of which can be configured to user preference as described below:
Clock
:
Set to
12-Hour
or
24-Hour
display preference. Then, set local time.
Temperature
:
Set to
Fahrenheit
or
Celsius
, as desired.
Standby Purge
:
When
ENABLE
selected, this function purges the chamber for 5
minutes every 2 hours. This occurs at regular intervals on every even hour, but only
when the Power Switch (see Fig. 6-3) is on. This function will not start if a program is
running. This purge reduces the build-up of CA fumes that accumulate inside the
chamber as CA deposits continually outgas. Selecting
NIGHT
will only enable this
function between 7pm and 5am.
Post Purge Time
:
This function keeps the blowers on after the Main Door is opened,
following the completion of a Fuming Cycle. This pulls lingering CA vapors away from
the user while removing evidence. There are three settings:
0
,
5
, or
10
minutes.
Selecting
0
will turn the blowers off when Main Door is opened, selecting
5
or
10
will
keep blowers on for 5 or 10 minutes after Main Door is opened. Closing Main Door
turns blowers off.
Mute
:
When
YES
selected, the keypad buttons will not beep when pushed.
Diagnostic
:
Allows internal component function to be tested for troubleshooting (see
Section 10: Troubleshooting
for more details).
Filter Reorder
:
Filter part numbers and Labconco Service phone number.
Figure 6-5
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...