Initial Chamber Set-Up
Once the CApture BT Fuming Chamber has been installed in its final location, the
following steps must be performed before running the chamber:
1. Remove tape from all doors and bottom drawer, and any remaining packaging
material from around the chamber. If not already completed.
2. Verify protective plastic is removed from Main Carbon Filter.
3. Remove the Perforated Floor from inside the main chamber, remove foam
sheeting from around Floor, then re-install Floor as shown in Fig. 6-2.
4. Remove the four (4) Hang Bars (shipped inside the Drawer) from their foam
sheeting, and place them as desired in the Side Wall Brackets (see Fig. 6-2).
Initial Chamber Cleaning
It is recommended to wipe down the inside of the main chamber with a dry cloth before
fuming evidence to ensure fingerprints from installation on the stainless steel walls or
inside of the glass door do not attract CA fumes. (Do
NOT
use chlorine-based solutions
to clean the unit!)
It is recommended to scrape the inside of the glass on the main door with a razor blade,
and wipe it with a dry cloth afterwards.
CA residue will deposit on the main glass door. After about 10 fuming cycles have been
completed, completely scrape the glass with a razor blade. This will significantly reduce
future build-up of CA deposits.
Figure 6-2
HANG BAR
PERFORATED FLOOR
SIDE WALL
BRACKETS
DRAWER
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...