40
CA Fuming
– Maintenance Notifications
If maintenance is required to the CApture BT Fuming Chamber, you will receive a
notification on the
ADD CA
screen before you begin a Fuming Cycle (see Fig. 6-11).
This is just a reminder notification, not an error or alert. Alerts are shown in the bottom
right corner, and a Fuming Cycle cannot begin if an Alert is shown. If a notification is
displayed, the Fuming Cycle will still begin when
[START]
is pressed.
The maintenance notification will be displayed in red text as shown above in Fig. 6-11.
See
Section 6: Maintaining Your CApture BT
for instructions on performing the
maintenance as notified.
The maintenance notification will be displayed based on the number of Fuming Cycles
that have been run in the unit. The frequency corresponds to the
Recommended
Maintenance Schedule
in Section 7. These maintenance operations may need to be
performed more frequently than what is listed in
Section 7
of this manual, or before the
maintenance notification is displayed on the screen. The frequency of maintenance is
based upon several factors, including, but not limited to, how much CA glue is used,
and how long the Fuming Cycles last. For more information on when to perform routine
maintenance, please see
Section 6: Maintaining Your CApture BT
.
If evidence is already loaded in the chamber when a maintenance notification is
displayed on the
ADD CA
screen, it is recommended to start the current cycle, and then
remove all evidence after the cycle is complete. Then, perform the maintenance after
the cycle, when the chamber is empty.
Figure 6-11
NOTIFICATIONS
DISPLAYED IN RED
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...