22
Remove the two Lift Brackets and the white protective spacer behind each bracket by
removing the two (2) bolts per bracket (requires ½” wrench or socket and ratchet).
Replace the bolts into the open threaded inserts and discard the Lift Brackets and
protective spacers if the chamber will not be moved again.
Chamber Installation
After the CApture BT Fuming Chamber has been removed from its pallet and positioned
in its final location, you must perform the following steps.
Step 1
1. Remove any tape securing doors, drawers and other items.
2. Retrieve the power cord from the white parts box.
3. Connect the power cord into the back of the chamber, and then into an
appropriately rated outlet.
Note: The following are located in the white parts box:
•
User’s Manual – on USB Thumbdrive (hard copy not included)
•
Power Cord
•
Hygrometer
•
Three (3) samples of Security Tags
•
Leg Levelers (4)
– not used if installed onto Accessory Stand
If you did not receive one or more of the components listed, or if any of the components
are damaged, contact Labconco Corporation immediately for further instructions.
Step 2
Warning:
The Main Carbon Filter is installed, but it remains sealed in a protective
plastic. This filter MUST be removed and the protective plastic discarded before the
chamber can be used. Failure to perform this procedure will result in CA fumes not
exhausting from the chamber. Follow these steps
1. Locate the warning sign on the Pre-Filter Cover. See Fig. 4-2. Loosen the single
thumbscrew on the front most flange of the Cover. Hinge the Pre-Filter Cover
open.
2. Locate and remove the two Phillips head screws on the front most flange of the
Main Filter Bracket. The Main Filter and Bracket will swing down about 20
degrees, but will stop on internal pins. Reference Fig. 7-5.
3. Slide the Main Carbon Filter out.
4. Remove the protective plastic cover.
5. Replace the Main Carbon Filter in its bracket (gasket on Filter goes up).
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...