38
Programs 1 through 3 offer three factory default options for common CA fuming
methods. These first three programs can be modified, and then run; however, the
program changes will not be saved. You can modify and save changes to Programs 4
through 20. The default parameters for Programs 1 through 3 are listed below in Table
5-1 along with a description for their use.
It is recommended to use a low viscosity Cyanoacrylate (very fluid like water) for best
results. Default Programs 1 and 2 are developed for use with a low viscosity
Cyanoacrylate. Arrowhead Forensics A-2601 CA was used to develop these first two
programs. Using a different CA, particularly a higher viscosity CA, may require
adjustments to the program parameters for best results.
Program 3 is developed for use with Lumicyano
TM
, which is a CA product with a dye that
fluoresces under a special laser and curved orange photography filter system. This
product can provide better contrast for developed fingerprints. Using a different CA
product with dye stain may require adjustments to the program parameters for best
results.
There are several factors that affect the outcome of the CA fuming process. These
preset Program options are just a helpful starting point for the end user. You will need
to test these setting and make adjustments as needed due to the specific glue, or
environment you are operating in.
Program #1
Program #2
Program #3
Grams CA
2.5
2.5
2.7
Humidity
80%
80%
75%
RH Dwell
0:00
2:00
0:00
CA Heat
350
O
F (177
O
C)
250
O
F (121
O
C)
250
O
F (121
O
C)
Fume
10:00
18:00
17:00
Purge
5:00
5:00
5:00
Description
New Prints, Faster
Run Time
Older Prints, More
Dwell Time
Dye Staining
(Lumicyano
TM
)
Table 6-1
Summary of Contents for Capture BT 31700-00
Page 2: ...2 CAptureTM BT Fuming Chamber 2020 Present 3170000 3170001 3170002 3170003 3170004 3170005...
Page 55: ...55 Figure 7 6...
Page 81: ...81 Wiring Diagrams 100 120V...
Page 82: ...82 208 230V...
Page 87: ...87 Chamber will not achieve humidity...
Page 88: ...88 Chamber does not exhaust CA fumes...
Page 89: ...89 Door alarm active when door s closed...
Page 90: ...90 Hot Plate will not achieve temperature set point...
Page 91: ...91 White CA residue builds on inside of Chamber...